Search results for "Ferrulized implant neck"

showing 1 items of 1 documents

In-vitro fatigue and fracture performance of three different ferrulized implant connections used in fixed prosthesis

2021

Background/purpose: The aim of the present in vitro study was to evaluate fatigue resistance of dental fixtures in three different types of fixture/abutment finishing line. Materials and methods: Transmucosal dental implants, with or without ferrulized neck, underwent fatigue tests (static and dynamic load) using the following standard protocol: UNI EN ISO 14801:2016. Two types of loading devices (screw- or cement-retained restoration) were also tested, and fatigue cycle tests were run to failure. Data of static and dynamic load tests were analyzed by proper statistical methods. Results: Following standard protocol for fatigue testing, the ILC type (Implant Level with ferrulized neck and ce…

OrthodonticsMaterials sciencemedicine.medical_treatmentFerruleDynamic load testingCrown (dentistry)lcsh:RK1-715Ferrulized implant necklcsh:DentistrymedicineFracture (geology)Ultimate failureOriginal ArticleImplantDeformation (engineering)Prosthesis fixationGeneral DentistryAbutment (dentistry)Fatigue testJournal of Dental Sciences
researchProduct